4.2.12 Output - DC Current
CALibration[:SOURce]:CURRent:DC[:HRANge][:FSCale]<NRf>
This command will result in the calibration coefficient being calculated for the DC full scale output
current in the high voltage range.
Command Syntax CALibration:CURRent:DC:HRANge <NRf>
Parameters <NRf> (value from 0 to 65535)
Examples CAL:CURR:DC:HRAN 35000
Query Syntax CALibration:CURRent:DCHRANge?
Returned Parameters <NRf> (value from 0 to 65535)
CALibration[:SOURce]:CURRent:DC[:HRANge]:ZERO <NRf+>
This command will set the calibration coefficient for the dc output current offset.
Command Syntax CALibration:CURRent:DC:HRANge:ZERO <NRf+>
Parameters <NRf> (a value between -2000 and +2000)
Examples CAL:CURR:DC:HRAN:ZERO -1003
Query Syntax CALibration:VOLTage:DC:HRANge:ZERO?
Returned Parameters <NR1> (value range -2000 to +2000)