Sequoia Series    75 
CURRent:HIGH 
This command sets the soft limits for maximum output current to which the unit could be programmed 
Command Syntax  SOURCE: CURRent: HIGH 
Parameters  <NR2> 
Unit  A (rms amperes) 
Examples  CURR:HIGH 15 
Query Syntax  CURR:HIGH? 
Returned Parameters  <NR2> 
Related Commands  OUTP:PROT:STATE  VOLT:RANG  MODE 
CURRent:LOW 
This command sets the soft limits for minimum output current to which the unit could be programmed 
Command Syntax  SOURCE: CURRent: LOW 
Parameters  <NR2> 
Unit  A (rms amperes) 
Examples  CURR:LOW 5 
Query Syntax  CURR:LOW? 
Returned Parameters  <NR2> 
Related Commands  OUTP:PROT:STATE  VOLT:RANG  MODE 
CURRent:ALC 
Phase Selectable 
This command enables or disables the automatic current level control mode.  If ALC is on, the output 
current of the power source is continuously adjusted based on the current measurement data acquired 
by  the  controller's  measurement  system.    This  results  in  increased  output  accuracy  and  load 
regulation. If regulation cannot be maintained, an output  current fault error is generated and the output 
is turned off. The regulate mode (REG) may be set to enable ALC mode but disable the output curr 
fault error if load regulation cannot be maintained. 
Note that  for the  Sequoia  1U, 1-Phase and Sequoia 2U, 3-Phase,  this mode  is coupled with the 
OUTP:IMP commands as the ALC mode and the programmable impedance functions are mutually 
exclusive. Attempting to enable one while the other is active will result in a command error. The ALC 
mode is also inactive during transient execution, even if it is enabled. 
Command Syntax  [SOURce:]CURRent:ALC <bool> 
Parameters  0 | OFF | 1 | ON | 2 | REG 
Unit  A/S (Ampere per second) 
*RST Value  0 
Examples  CURR:ALC 1 
Query Syntax  [SOURce:]CURRent:ALC? 
Returned Parameters  0 | 1 | 2 
Related Commands  OUTP:IMP:STAT