Section 5 Remote Control
5-20
(2) Extended ERR event status register
Extended ERR event status enable register
Extended ERR event status register
7
6
Logical OR
5
4
3
2
1
0
7
6
5
4
3
2
1
0
&
&
&
&
&
&
&
&
Status byte register
Set to ERR bit
FPGA configuration error
Parameter error
EEPROM write error
Not used
Not used
Not used
Not used
Not used
Bit Event name Description
7 Not used Always 0.
6 Not used Always 0.
5 Not used Always 0.
4 Not used Always 0.
3 Not used Always 0.
2 EEPROM write error Set when writing to EEPROM failed.
1 Parameter error
Set when parameters cannot be restored by automatic pa-
rameter restoration at power-on.
0
FPGA configuration er-
ror
Set when FPGA configuration failed during self-test at
power-on.