A3 PLUS - A5 - A6 PLUS - L6
SURGICAL SINGLE CART
12. DENTAL UNIT MAINTENANCE SCHEDULE
Eliminate condensate from air ducts.
2-minute flushing of the ducts of the instruments used
on the patient (Manual or LONG FLUSHING)
Air-liquid separator canister by
CATTANI
Insert inside each suction filter a defoaming tablet.
Disposable infection control sheaths.
Apply flexible sheaths and disposable barriers on
dental chair and dental unit.
Lubricate turbine and micromotor handpieces.
See documentation
enclosed with instrument.
W.H.E. system
(if the dental unit was left off for more
than 7 days)
Empty the W.H.E. system water circuit
20-second flushing of the instruments used on the
patient (Manual or QUICK FLUSHING)
See documentation
enclosed with instrument.
Carry out a STANDARD flushing cycle (if any) or suck
in about half a litre of 6% diluted STER 3 PLUS
solution with each of the suction tubes used.
Sterilize the contaminated flexible protections.
Replace disposable barriers.
Clean surfaces with STER 1 PLUS.
Disinfection of spray water ducts (if the BIOSTER
system is fitted).
Clean filter in running water.
The content must be disposed of separately.
Clean with commercially-available detergents specific
for the material composing it.
Do not use acids or abrasive products.
Pour approximately 1 litre of STER 3 PLUS diluted in
a 6% solution,
Wash with a specially formulated scale-remover.
Check filter and, in case suction capacity is reduced,
change it (code 97461845)
Carry out an automatic flushing cycle (if any) or suck
in about half a litre of 6% diluted STER 3 PLUS
solution with each of the suction tubes used.
Suction tube holder terminals.
Sterilize in water steam autoclave.
Hydraulic saliva ejector.
Clean the filter of the saliva ejector terminal.
Sterilize the present flexible protections.
Eliminate all used disposable barriers.
Dental unit and chair surfaces.
Clean surfaces with STER 1 PLUS.