A3 PLUS - A5 - A6 PLUS - L6
SURGICAL SINGLE CART
9.3. SUCTION FILTER CLEANING
This operation should be done daily at the end of work.
Always wear goggles and gloves to prevent contact with
infected material when cleaning the suction filters.
Dental units of A3 PLUS model:
• Remove the filter (d).
• Clean/replace the filter (code 97461845).
• Refit the filter (d).
Before refitting the filter, make sure to remove any
amalgam residues still present on filter seat mouth.
In order to prevent any dripping of liquids and secretions from
the filter being removed, it is advisable to carry out the above
operations with the suction tube working.
Dental units of A5 and A6 PLUS model:
• Remove filter door (e) from its seat by pulling it up.
• One at a time, remove filters (d).
The filter door can be used as a container for the transfer to
the cleaning area.
• Clean/replace the filters (code 97461845).
• Refit filters.
Before refitting the filter, make sure to remove any
amalgam residues still present on each filter seat mouth.
In order to prevent any dripping of liquids and secretions from
the filter being removed, it is advisable to carry out the above
operations with the suction tube working.
Dental units of L6 model:
• Remove filter door (e) from its seat by pulling it up.
• One at a time, remove filters (d).
The filter door can be used as a container for the transfer to
the cleaning area.
• Clean/replace the filters (code 97461845).
• Refit filters.
Before refitting the filter, make sure to remove any
amalgam residues still present on filter seat mouth.
In order to prevent any dripping of liquids and secretions from
the filter being removed, it is advisable to carry out the above
operations with the suction tube working.