A3 PLUS - A5 - A6 PLUS - L6
SURGICAL SINGLE CART
4.2. MOVEMENT LOCK DEVICES
Use the devices below when movement of the equipment
needs to be blocked:
• Dental chair control buttons (a) or (c).
By pressing any of the dental chair control buttons all
movements will be stopped.
• Foot control (b).
When the foot control is actuated, all movements of the
equipment are blocked.
• Stop foot board (l).
When the floor board is actuated, all movements of the
equipment susceptible to cause squashing are blocked.
4.3. ADJUSTABLE HEADREST
Warnings for use.
• Do not attempt to move headrest while patient is resting against it.
• Do not attempt to modify the position of the cushion without first releasing the lock mechanism.
• The pneumatic locking device is active only when the air circuit is pressurised and the dental unit is on.
Headrest models
Available models are:
1 with manual cushion retainer
2 with air-operated cushion retainer
Adjusting the manual headrest
• Lift or lower the headrest to the desired position.
• Unlock the cushion by turning the lock knob (k) counter-clockwise.
• Aim the cushion as desired.
• Lock the cushion again by turning the knob (k) clockwise.
Adjusting the air-operated headrest
Only when the dental unit is on.
• Raise or lower the headrest while holding down button (u).
• Aim the cushion as desired while holding down button (u).
Proper positioning of the headrest.
For correct use of the headrest, position the patient's head
as shown in the figure.