A3 PLUS - A5 - A6 PLUS - L6
SURGICAL SINGLE CART
EN
6.4. TRAY HOLDER
The tray holder (a) is made of stainless steel and can easily be removed
from its support.
Tray holder support can rotate both clockwise and counter-clockwise, so
as to be positioned in the most convenient position for the operator.
To lock/unlock the tray holder support, simply turn the clutch knob (b).
Maximum permitted load on the tray holder: 1 kg
distributed.
6.5. HYDRAULIC SALIVA EJECTOR
Warnings for use.
The dental unit equipped with hydraulic saliva ejector does
not comply with standard EN 1717.
The hydraulic saliva ejector starts running when the tube is removed
from the support.
Cleaning after each use.
Aspirate about ½ litre of STER 3 PLUS (CEFLA s.c.) diluted in a 6%
solution (equivalent to 60 ml of product in 1 litre of hot water).
STER 3 PLUS must be used as specified by the
Manufacturer.
Cleaning the saliva ejector filter.
This operation must be carried out at the end of each work day.
Wear gloves before attempting to perform this operation!
Aspirate about ½ litre of STER 3 PLUS diluted in a 6% solution
(equivalent to 60 ml of product in 1 litre of hot water).
In order to prevent possible dripping of liquids and matter from the filter
(b) to be extracted, suck only air for about 5 seconds.
Take off the cap (a) by turning and pulling at the same time.
Remove the filter (b).
Clean/replace the filter (code 97290060).
Put the filter and cap back in place.
To prevent liquids and matter from dripping from the filter
taken out, suck air only for approximately 5 seconds before
performing these operations.
Routine maintenance.
Lubricate the O-Ring (c) using S1-Protective Lubricant for O-Rings.