A3 PLUS - A5 - A6 PLUS - L6
SURGICAL SINGLE CART
9.6. CLEANING THE TURBINE RETURN AIR FILTER
Monthly check the oil container filter (g) present in the turbine’s return air line.
If necessary, replace the filter element (code 97290014).
If the micromotor instrument is also present on the assistant's board, remember that the relevant filters must be cleaned.
9.7. METASYS AMALGAM SEPARATOR
The maintenance instructions for the METASYS amalgam separator are enclosed with the equipment, if the equipment comes with this type of
separator.
The separator control device is located in the water unit.
• Always wear goggles and gloves to prevent contact with infected material when cleaning the separator.
• When disposing of disposable containers full of amalgam, observe current local and national laws.