A3 PLUS - A5 - A6 PLUS - L6
SURGICAL SINGLE CART
The suction lines must be disinfected using a product suitable for this
purpose.
For cleaning of the suction system, it is recommended to
use STER 3 PLUS (CEFLA s.c.) diluted in a 6% solution
(equivalent to 60 ml of product in 1 litre of hot water).
STER 3 PLUS must be used as specified by the
Manufacturer.
At the end of each surgical procedure.
• Perform an automatic flushing cycle or suck in about 1/2 litre of
sanitising solution with each one of the suction tubes used.
• Sterilize the suction tube holder terminals in a steam autoclave
(see paragraph 1.6).
At the end of each working day.
• Suck in 1 litre of water with each suction tube, alternating water and air
(keep the suction tube alternately in and out of the water).
• After rinsing with water, suck in about 1/2 litre of sanitising solution
with each one of the suction tubes used.
All sanitising products must be used as specified by the
manufacturer.
After flushing and disinfection, it is good practice to suck in air
only to dry the entire suction system (5 minutes).
Once a week.
Remove the suction tube body from its hose attachment and lubricate
the O-rings (o) using S1-Protective Lubricant for O-Rings (CEFLA s.c.).
Once a year.
Replace the suction tubes and suction tube holder terminals.