A3 PLUS - A5 - A6 PLUS - L6
SURGICAL SINGLE CART
5.2.3. "POWER PEDAL" FOOT CONTROL
Description of the parts.
1 Handle.
2 Foot control.
3 Dental chair movement control.
4 Chip-air control or activation/deactivation of instrument spray
function.
5 Water Clean System control or activation/deactivation of
instrument spray function.
6 Spray operation LED.
7 Battery charge LED (wireless version only).
8 Automatic dental chair return or programme "B” recall activation.
9 Patient rinse position or programme "A" recall activation.
Foot control (2).
With instrument removed
• Pushing the pedal (2) for a long time starts the instrument.
You can adjust the speed/power by modulating the pressure on the
pedal.
The foot control adjusts the speed/power of the instrument
from the minimum to maximum value set from the dentist's
board.
• Release the pedal (2) to stop instrument operation.
With the spray active, at the end of the operation a blast of air
is automatically activated to remove any residual drops of
liquid in the spray ducts.
With instruments at rest
• When the foot control (2) is pressed, all automatic dental chair
movements are automatically blocked.
Joystick for dental chair movements (3).
It controls the following movements:
A Dental chair seat up.
B Dental chair backrest up.
C Dental chair seat down.
D Dental chair backrest down.
To stop movement, release the control.
• All dental chair movements are blocked when an instrument
is being used or the BIOSTER system is running.
The setting made by the Technical Service allows changing the joystick
operation when the instruments are extracted but not active:
A Operating light ON/OFF control.
B Micromotor rotation direction reversal ON/OFF control, scaler
ENDO function activation, camera MIRROR function enabling.
C Instrument memory change.
D Activation from FLUO LED to WHITE LED, and vice versa, with
micromotor in RESTORATIVE mode.
Peristaltic pump activation ON/OFF control in the other modes.