SURGICAL SINGLE CART
EN
27
Cleaning the dentist's board.
Clean the dentist's board using a suitable product (see paragraph 1.4).
x Removable instrument holder; just slide it out of its seat to remove it.
m Board handle that can be removed and sterilized in an autoclave at up
to 121°C.
To remove the handle, first press the special locking buttons.
v Silicone protection that can be sterilized in an autoclave at up to
135°C.
Perform cleaning and disinfection operations after each
patient.
Removable instrument hoses.
Clean and sanitise the external surface of the instrument hose using a
suitable product (see paragraph 1.4).
Instrument hoses are NOT designed for sterilization in
autoclave or by being cold soaked in solution.
Boards, version with Upper Recovery: to remove hoses, first
remove instrument holder (x).
Before removing instrument hoses, turn dental unit off.
After having turned dental unit off, drain syringe ducts by pressing the relevant air and water buttons directly on bowl until all
spray water has flowed out.
TURBINE, MICROMOTOR and SCALER instrument hoses contain water; it is therefore recommended to remove the hose by
keeping the handpiece side end on the bowl.
When refitting a hose, make sure that electric contacts are dry and that the plastic ring nut is well tightened.
Each hose must be refitted only and exclusively inside the corresponding instrument holder.
Fixed instrument hoses.
Clean and sanitise the external surface of the instrument hose using a suitable product (see paragraph 1.4).
To remove hose, contact the Technical Service.