A3 PLUS - A5 - A6 PLUS - L6
SURGICAL SINGLE CART
Cleaning the dentist's board.
Clean the dentist's board using a suitable product (see paragraph 1.4).
x Removable instrument holder; just slide it out of its seat to remove it.
m Board handle that can be removed and sterilized in an autoclave at up
to 121°C.
To remove the handle, first press the special locking buttons.
v Silicone protection that can be sterilized in an autoclave at up to
135°C.
Perform cleaning and disinfection operations after each
patient.
Removable instrument hoses.
Clean and sanitise the external surface of the instrument hose using a
suitable product (see paragraph 1.4).
Instrument hoses are NOT designed for sterilization in
autoclave or by being cold soaked in solution.
Boards, version with Upper Recovery: to remove hoses, first
remove instrument holder (x).
• Before removing instrument hoses, turn dental unit off.
• After having turned dental unit off, drain syringe ducts by pressing the relevant air and water buttons directly on bowl until all
spray water has flowed out.
• TURBINE, MICROMOTOR and SCALER instrument hoses contain water; it is therefore recommended to remove the hose by
keeping the handpiece side end on the bowl.
• When refitting a hose, make sure that electric contacts are dry and that the plastic ring nut is well tightened.
• Each hose must be refitted only and exclusively inside the corresponding instrument holder.
Fixed instrument hoses.
Clean and sanitise the external surface of the instrument hose using a suitable product (see paragraph 1.4).
To remove hose, contact the Technical Service.