SURGICAL SINGLE CART
Selecting the micromotor working programmes (LCD TOUCH
console only).
It is possible to set 4 micromotor working programmes.
Each working programme saves the following data:
• Operating mode.
• Maximum rotation speed.
• Handpiece reduction ratio.
The data set are automatically saved.
• Press button (E) repeatedly to set the desired working programme.
Selecting the reduction ratio (LCD TOUCH console only).
Select the desired reduction ratio using buttons (C).
The data set are automatically saved.
Activation/deactivation of independent water supply.
See paragraph 7.2.
Removable hose.
The micromotor features a removable hose to make cleaning operations easier (see paragraph 5.).
Cleaning and maintenance.
Refer to the specific instructions provided with the instrument.
It is recommended to use Daily Oil Plus (CEFLA s.c.) for lubrication.
• Do not soak the instrument in disinfectant fluids or detergents.
• Inadvisable products: abrasive products and/or products containing acetone, chlorine and sodium hypochlorite.
Sterilization.
Instrument handpiece only: see paragraph 1.6.
Carefully read the operating instructions supplied with the instrument before attempting to sterilize.
• The instrument is supplied NOT sterile and must be sterilized before use.
Carefully read the operating instructions supplied with the instrument before attempting to sterilize.
• Never put the contra angle on the micromotor while it is running.
• The drill release button must not be pressed during operation!
Friction between the button and micromotor rotor overheats the head and may cause burns.
• The patient’s internal tissues (tongue, cheeks, lips, etc.) must be protected against contact with the button by using suitable
instruments (mirrors, etc.).
• The drills and various instruments attached to the handpieces must comply with Biocompatibility Standard ISO 10993.