SURGICAL SINGLE CART
EN
Fibre optic brightness adjustment.
Place the instrument in its work position.
To adjust fibre optic brightness, press the button (F) (for at least 2
seconds).
Adjust beam brightness by pressing “increase” or “decrease” buttons.
Value setting range: 1 to 16.
To confirm set brightness, press the button (F) (for at least 2
seconds).
Scaler working programme selection
(for LCD Touch DISPLAY only).
The scaler has 3 different working programmes, identified with P1, P2
and P3, that can be selected cyclically.
Each working programme saves the following data:
maximum power,
fibre optic on/off,
type of delivered spray.
Touch the button (A) repeatedly to select the desired programme.
The variation is cyclical.
Scaler operating mode selection
(for LCD Touch DISPLAY only).
With instrument removed
Select the scaler operating mode by touching the button (D)
repeatedly, the following operating modes will be selected in
sequence:
D1 NORM (normal).
D2 ENDO (50% reduced power).
D3 PARO (40% reduced power).
With the foot control lever activated, you cannot edit operating
mode.
The data set are automatically saved.
Scaler operating mode selection
(DIGIT console).
With instrument removed
Select the scaler operating mode by touching the button (E)
repeatedly, the following operating modes will be selected in
sequence:
NORM (normal).
ENDO (50% reduced power).
The activation of ENDO mode is signalled on the display
through icon (C).
With the foot control lever activated, you cannot edit operating
mode.
The data set are automatically saved.
Removable hose.
Scaler features a removable hose to make cleaning operations easier (see paragraph 5).
Cleaning and maintenance.
Refer to the specific instructions provided with the instrument.
Do not soak the handpiece in disinfectant liquids or detergents.
Sterilization.
Torque wrench, scaler bits and scaler handpiece: steam autoclave up to 135°C following the instructions for the use of the device
Carefully read the operating instructions supplied with the instrument before attempting to sterilize.