A3 PLUS - A5 - A6 PLUS - L6
SURGICAL SINGLE CART
Spout, bowl and bowl filter removal.
• Pull up the spout (l) and take it off.
• Pull up the filter (q) and its cover (p) from the bowl to remove them.
• Release bowl (m) by turning it counter-clockwise, then pull it up to
remove it.
Disinfection and cleaning.
Always wear gloves to prevent contact with infected
material when cleaning the bowl and bowl filter.
The parts are to be cleaned daily at the end of each working day.
• Water dispensers: thoroughly wash with a specially formulated scale-
remover.
• Bowl filter: clean with running water and commercially-available
cleaning products.
• Bowl: clean with commercially-available detergents specific for the
material composing it.
• Bowl drain line: pour 1 litre of STER 3 PLUS diluted in a 6% solution
(equivalent to 60 ml of product in 1 litre of hot water), this operation
improves drain line efficiency and prevents blockages.
Do not use acids or abrasive products.
STER 3 PLUS must be used as specified by the
Manufacturer.
Resin bowl: it can undergo thermal disinfection at 90° for 5
minutes (max. 50 cycles).