A3 PLUS - A5 - A6 PLUS - L6
SURGICAL SINGLE CART
Setting the disinfection cycle.
• Check disinfectant fluid level inside tank and, if necessary, top up.
The disinfection cycle will not be activated if the fluid level
inside tank is below reserve level.
• Enter the BIOSTER disinfection cycle by pressing and holding button
(H1) on the assistant's board.
LCD TOUCH console only: disinfectant fluid residence time
in the water ducts can be set (see paragraph 5.1.1.1.3.).
• Dental units of A3 PLUS, A5 and A6 PLUS model: position the special
container (d) on the bowl and insert the instrument hoses to be
disinfected.
Dental units of L6 model: remove the cover (b).
Press the edge (b1) of the cover to lift it.
• Insert the hoses of the instruments to be disinfected in the suitable
seats.
For the syringe instrument, it is necessary to remove the
handle and completely insert the dedicated adapter (f)
(second click). The heating system must be switched off.
The micromotor hose must be fully inserted inside motor
body.
Turbine and scaler hoses must be inserted without the
handpiece.
• Should you wish to disinfect suction tubes, insert tube terminals inside
the special couplings under manifold (see paragraph 7.5.).
Make sure that suction tube terminals are open.
• If water-to-cup duct disinfection is selected, place the special supplied
cup (e) under cup spout.
Make sure that spray cocks (g), positioned on board lower side, are fully
open.
Performing the disinfection cycle.
• Start the automatic disinfection cycle by touching the button (A) on the
LCD TOUCH display or by pressing the button (H1) on the assistant's
board.
• Now the system automatically performs the following operations:
- Emptying water ducts of instruments with air.
- Introduction of disinfectant fluid and starting of the counting of the
previously-set residence time.
- Once time has elapsed, start of a new duct emptying phase with air.
- Ducts flushing with mains water or with independent supply fluid
(only if S.H.S. system is installed and active).
• Disinfection cycle time: approx. 20 minutes (depending on the number
of instruments being disinfected).
Disinfection cycle interruption.
• Touch the “PAUSE” button (A1) to interrupt the disinfection cycle at
any time.
• Touch the “STOP” button (A2) within 5 seconds to cancel the
disinfection cycle definitively.
• After 5 seconds, touch the “PAUSE” button (A1) to resume the
disinfection cycle from where it was interrupted.
DIGIT console only: once the disinfection cycle has started, it
can no longer be interrupted.
PEROXY Ag+ storage.
For proper PEROXY Ag+ storage, follow the manufacturer’s instructions given on the package.
It is important to keep the package tightly closed and stored in a cool place at a temperature not exceeding 25°C.
Never leave PEROXY Ag+ or oxygenated water in the tank (a) for more than one month.
After your absence from the surgery for long periods of time (holidays), completely empty the tank (a) before restarting the dental
unit.