HA-8180V 3-37
For sample cups:
Odd-numbered ports: Empty
Even-numbered ports: Sample cup
e Load the hemolysis control rack onto the sampler.
2 Measure the samples.
a Press .
Measurement will start.
The obtained result will appear on the display and a
chromatogram report will be printed after each measurement.
If “Measurement ended.” appears:
The most recent result will appear on the display.
If an error or trouble message appears:
Trouble was detected. See the relevant page in “Chapter 5
Troubleshooting” to take the appropriate action.
b To return to the standby screen, press four times.
c Check that the hemolysis control rack is not moving, and remove
the rack from the rack unloading side of the sampler.
Sample cup with
sample
Empty
12345678910
Hemolysis sample measurement
0"40
(8520) 15.0 kg/cm2
Measuring ...
Meas count 1 time(s)
HbA1c HbF
Meas value
35 0.7
Mean value
35.0 0.70
R
0 0.0
S.D.
--.- --.--
C.V.
--.-
Stop
REFERENCE:
To stop measurement, press or .
Start
Stop
Hemolysis sample measurement
(8520) 10-06-2010 10:11
Measurement ended.
Meas count 10 time(s)
HbA1c HbF
Meas value
35 0.7
Mean value
35.1 0.68
R
1 0.1
S.D.
0.3 0.04
C.V.
0.8
Go back
Go back