HA-8180V 4-25
c Fit the spare tubes into the rear hooks.
d Wipe up if there is any leaked liquid.
e Close the side cover.
f Place the hemolysis washing solution bottle in its original
position.
Connect tubes to the correct joints. Liquid may leak if
tubes are connected to the wrong joints or loose. The
instrument may be damaged as a result.
Check tubes are not kinked at the joints. If the tubes are
kinked, it may cause liquid waste to obstruct the flow.
There are two hooks on each valve. Be sure to pinch the
tube in the REAR hook with an arrow mark.
You can record the date of tube replacement on the instrument.
See “4.5.1. Recording the Date of Maintenance” on page 4-42.
4.2 Replacement of Consumables