HA-8180V 4-25
c Fit the spare tubes into the rear hooks.
d Wipe up if there is any leaked liquid.
e Close the side cover.
f Place the hemolysis washing solution bottle in its original
position.
NOTE:
Connect tubes to the correct joints. Liquid may leak if
tubes are connected to the wrong joints or loose. The
instrument may be damaged as a result.
Check tubes are not kinked at the joints. If the tubes are
kinked, it may cause liquid waste to obstruct the flow.
Tube
Seen from above
NOTE:
There are two hooks on each valve. Be sure to pinch the
tube in the REAR hook with an arrow mark.
REFERENCE:
You can record the date of tube replacement on the instrument.
See “4.5.1. Recording the Date of Maintenance” on page 4-42.