HA-8180V 1-27
For the bottle for liquid waste (other than the optional
product) or lab drainage system
Insert the tubes into the bottle for liquid waste or lab drainage
system. Make sure tubes are long enough when cutting, to
prevent them from being pulled from the bottle or drainage
system.
For the optional liquid waste bottle
Connect the tubes to the cap of the liquid waste bottle. Make sure
tubes are long enough when cutting, to prevent the bottle from
falling over.
3 Connect the fluid level detection sensor cord.
a When using the optional liquid waste bottle, connect the fluid
level detection sensor cord to the DRAIN terminal on the rear
panel.
Go to the following sections as needed:
• “1.4.7. Connecting Peripheral Devices” on page 1-28.
• “1.4.8. Connecting the Power Cord” on page 1-29.
Keep the remaining tubes in the accessory case.
1.4 Installation