HA-8180V 2-36
Chapter 2 Measurement
7 Start control measurement.
a On the standby screen, check the [Measurement mode] setting.
To change the mode, see “2.4. Measurement Mode Selection (Variant or Fast)” on page 2-14.
b On the standby screen, press .
“Preparing for measurement” will appear and the sample rack
will move to the aspiration position.
“Measuring ...”
One to three measurements will be performed per sample
container, as selected in the <Control measurement count>
setting (see “3.3.5. Setting the HbA1
C Measurement
Conditions” on page 3-13).
Control will not be stirred.
“Results”
The obtained results will be displayed on the screen and
printed out.
<A1c Cont> shows the control measurement number.
See “2.9. Displayed and Printed Reports” on page 2-44.
8 Once the control measurements are complete
a Check that the sample racks are not moving, and remove the sample racks from the rack unloading side of the
sampler.
IMPORTANT:
Measurement accuracy is controlled separately for the Variant mode and Fast mode.
Accuracy control reports can be printed for each mode or both modes (see “3.7.3. Printing Accuracy Control
Reports” on page 3-31).
REFERENCE:
When the [Confirming HbA1c calibration] screen appears:
See “2.4. Measurement Mode Selection (Variant or Fast)” on page 2-14.
Measuring ...
0"45
15.0 kg/cm2
Sampling in process
A1c Cont 0030 Port No. 0011
ID 012345678901234569
Measurement : Variant
Column 0000
ABCH
Menu Stop Cal
REFERENCE:
Control measurement results can be tabulated everyday and used to check trends in measurement results. See
“3.7.3. Printing Accuracy Control Reports” on page 3-31.