EasyManua.ls Logo

Astar PhysioGo.Lite Electro - Page 12

Astar PhysioGo.Lite Electro
96 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4. Operational safety
User ManualPhysioGo.Lite Electro page 12 / 96
Immediately disconnect the patient in the case of appearing warning on error messages on the display.
Sitting or reclining position should be applied to the patients with respiratory disorders or breathing
difficulties.
Treatment parameters and electrodes placement should be consistent with the medical indications.
Connect the electrodes to the patient at a time when the device does not generate the electricity to avoid
the risk of electric shock.
Electrodes should not be placed alongside the carotid artery (carotid sinus), in the area of reproductive
organs, in the lower abdomen and over the internal organs.
It is prohibited to leave the patients unattended during treatments of electrotherapy.
It is necessary to take precautions in case of the occurrence of surface metal implants in the spot of
application.
Do not use electrotherapy at the area of application of surgical staples in the skin, or on tissues protected
with dressings or materials containing metal ions (silver, zinc).
If it is possible, the treatment polarity should be adjusted so that the negative pole ought to be ”further”
from the heart than the positive one.
It is not recommended to place electrodes in chest area, as it may increase the risk of ventricular
fibrillation.
Do not place electrodes on the neck and transcranially for epileptic patients, because stimulation may
cause seizures.
Unless specifically indicated by a doctor, avoid placing electrodes that form the circuit on the chest and
upper back or crossing over the heart.
Unless specifically indicated by a doctor, avoid applying electrical stimulation directly on the eyes or
mouth.
In case of treatment performed near the head, the patient should be in lying position.
Simultaneous performance of electrotherapy treatments and therapies with the use of high frequency
equipment (diathermy and electro surgery) may result in burns where electrodes are applied.
Take special care with patients with disturbed superficial sensation.
It is necessary to use operational and sanitized electrodes. Inadequate choice of electrodes may cause
skin irritations or burns.
It is recommended to differentiate the electrodes size according to performed treatment in order to do
not exceed the current density:
0,2 mA/cm
2
for currents with constant component (unipolar) galvanic, diadynamic, pulse currents,
unipolar sine surge, tonolysis,
2 mA/cm
2
for bipolar currentsTENS, Kotz, interferential.
Improper selection of electrodes can cause skin irritation and burns.
Carrying out treatments, where for the applied electrodes it is necessary to set the current/voltage so
that the current density does not exceed 2 mA/cm
2
, may require particular attention of the
physiotherapist.
Irritation and skin burns may occur during performing electrical stimulation. If such syndromes occur you
are obliged to interrupt the treatment and immediately consult with a doctor.
It is not recommended to apply unidirectional currents in CV mode due to the possibility of skin burns. It
is necessary to apply CC mode.
Unless specifically indicated by a doctor, it is not recommended to treat pregnant women with
electrotherapy.
Special caution must be kept during electrotherapy treatments in children and older people.
Battery use (optional):
Rechargeable battery UR18650A 5S1P is designed for use only with Astar PhysioGo.Lite range of devices.
In case of mechanical damage to the battery module, there is a risk of fire, explosion or burns, due to
used lithium-ion cells.
Do not throw the battery or hit it. Do not heat it or incinerate.
Do not short-circuit the contacts or disassemble the casing.
Do not immerse in liquids.
Operation, storage and transport conditions are presented in chapter 4.2.

Table of Contents

Related product manuals