8. Definitions and parameters
User ManualPhysioGo.Lite Electro page 46 / 96
Parameters description:
Symbol Description
Symbol on the
treatment
screen
Available parameters
Treatment time - 1 60 minutes, 1 minute step
Shape of the
current
SYM
Symmetric
ASYM
Asymmetric
ALT
Alternating asymmetric
TENS pulse
duration
t_imp
Possible settings: 25 µs, 50 µs, 75 µs, 100 µs, 150 µs, 200
µs, 250 µs, 300 µs, 400 µs, 500 µs
Note: pulses longer than 200µs are available only for the
sum of the frequencies (basic and spectrum) < 150Hz
Basic frequency
f_bas Regulation in the range of 1 Hz 200 Hz (variable step)
Frequency
spectrum
f_spec Regulation in the range of 0 Hz 200 Hz (variable step)
Frequency
modulation
program
FM
Frequency modulation program is switched
off
Frequency rise time 1 s
Hold time of maximum frequency 2 s
Frequency fall time 1 s
Hold time of basic frequency 3 s
Frequency rise time 3 s
Frequency fall time 3 s
Frequency rise time 6 s
Frequency fall time 6 s
Frequency rise time 6 s
Hold time of maximum frequency 6 s
Frequency fall time 6 s
Hold time of basic frequency 6 s
Frequency rise time 12 s
Frequency fall time 12 s
Random pulse generation
Amplitude
modulation
program
AM
Amplitude modulation program is switched
off
Amplitude rise time 3 s
Hold time of maximum amplitude 3 s
Amplitude fall time 3 s
Amplitude rise time 6 s
Amplitude fall time 6 s
Amplitude rise time 6 s
Hold time of maximum amplitude 12 s
Amplitude fall time 6 s
BURST mode FM = BURST
7 pulse sequences, 2 Hz
7 pulse sequences, 4 Hz
9 pulse sequences, 2 Hz
9 pulse sequences, 4 Hz
Amplifier
operation mode
-
CC stabilized output current
CV stabilized output voltage