8. Definitions and parameters
User ManualPhysioGo.Lite Electro page 60 / 96
Symbol Description
Symbol on
the treatment
screen
Available parameters
Pause duration t_pause
Max. regulation in the range of 1 ms – 5 s (variable step,
range depends on pulse duration)
Regulation in the range of 1 ms 200 ms for the
operation of training program PT (variable step)
Basic pulse
frequency
f_bas
calculated from the formula f_bas=1 / (t_imp + t_pause)
changing the parameter means a change of t_pause
Training
program
PT
No training program
Rise time 1 s
Contraction phase 2 s
Fall time 1 s
Rest phase 2 s
Rise time 1 s
Contraction phase 3 s
Fall time 1 s
Rest phase 8 s
Rise time 1 s
Contraction phase 5 s
Fall time 1 s
Rest phase 5 s
Rise time 1 s
Contraction phase 5 s
Fall time 1 s
Rest phase 10 s
Rise time 2 s
Contraction phase 10 s
Fall time 2 s
Rest phase 10 s
Rise time 2 s
Contraction phase 10 s
Fall time 2 s
Rest phase 20 s
Rise time 2 s
Contraction phase 10 s
Fall time 2 s
Rest phase 30 s
Rise time 2 s
Contraction phase 10 s
Fall time 2 s
Rest phase 40 s
Rise time 2 s
Contraction phase 10 s
Fall time 2 s
Rest phase 50 s
Rise time 3 s
Contraction phase 15 s
Fall time 3 s
Rest phase 30 s
Rise time 3 s
Contraction phase 15 s
Fall time 3 s
Rest phase 50 s