8. Definitions and parameters
User Manual – PhysioGo.Lite Electro page 74 / 96
Symbol Description
the treatment
Available parameters
Maximum amplitude -
0,3 – 140 mA for unipolar trigger and stimulating
pulses (channel A and B)
Regulation step:
• 0,2 mA in the range of 0,3-0,5 mA
• 0,5 mA in the range of 0,5-140 mA
0,5 – 140 mA for bipolar stimulating pulses (channel B)
Regulation step:
• 0,5 mA in the range of 0,5-140 mA
CV mode:
1,5 – 100 V, max. 100 mA for trigger pulses (channel A)
and bipolar stimulating pulses (channel B)
Regulation step:
• 0,5 V in the range of 1,5-100 V
1,5 – 100 V, max. 80 mA for unipolar stimulating pulses
(channel B)
Regulation step:
•
0,5 V in the range of 1,5-100 V