11. Specification and accessories
User Manual – PhysioGo.Lite Electro page 90 / 96
Immunity test IEC60601 test level Compliance level
Magnetic field power frequency
Hz)
IEC 61000-4-8
30 A/m 30 A/m
Immunity test IEC60601 test level Compliance level
Voltage dips
IEC 61000-4-11
0% U
T
0,5 cycle, phase angles of synchronization with AC power supply
voltage 0°, 45°, 90°, 135°, 180°, 225°, 270°, 315°
Complies
0% U
T
1 cycle,
phase angle of synchronization with AC power supply
voltage 0°
Complies
70% U
T
25 cycles for 50 Hz
30 cycles for 60 Hz
phase angle of synchronization with AC power supply voltage 0°
Complies
Voltage interruptions
IEC 61000-4-11
0% U
T
250 cycles for 50 Hz
300 cycles for 60 Hz
Complies
Immunity test Compliance level
Proximity fields from RF wireless communications equipment according to 8.10 IEC
60601-1-2:2014
Complies