EasyManua.ls Logo

Bauer Kompressoren Verticus 5 - Fig. 78 Hex Switch, Operating Mode Position

Bauer Kompressoren Verticus 5
168 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Fig. 76 Display Setup
Fig. 77 Calibrating the Touch Panel
Fig. 78 Hex switch, operating mode position
1
Instruction Manual w VERTICUS 5 Compressor Units
A-63
Display Setup
Contrast and brightness of the display are set on the Display
Setup page (Fig. 76).
Touch calibration
If the touch display is not working properly recalibration
may produce relief.
To do so, press the "Touch calib.” box on the display setup
page (Fig. 76). An new window containing a cross-hair is
displayed. Touch the calibration cross-hair (Fig. 77). Ano
ther cross appears. Touch this again and so on until all 4 cor
ners have been calibrated.
If the calibration fails (by touching the wrong position), the
calibration menu cannot be called up again. In this case, de-
energise the display and set the bottom hex switch (x16,
see 1, Fig. 78) on the back of the display from 0 to 2. Then
boot the unit and carry out the calibration again in the
calibration menu that appears. Then set the hex switch
back to 0 (as shown in Fig. 78), so that the display shows the
home page after rebooting (in position 2 the calibration
menu would rappear again).

Table of Contents

Related product manuals