Waste container components
The following figure shows the main components of the waste container.
5
4
3
21
Item Number Component
1 Baffle
2 User-replacable cap
3 Air line
4 Alarm sensor
5 Waste fluid line with clip
Contact with biological specimens and materials can transmit potentially fatal disease.
To prevent exposure to biohazardous agents:
l
Put the flow cytometer in standby mode before disconnecting the waste tank, to avoid leakage of
biohazardous waste.
l
Always disconnect the waste tank from the flow cytometer before you empty it. Wait at least 30seconds for
pressure to dissipate before you remove the waste cap or sensor.
l
Expose waste tank contents to bleach (10% of total volume)for 30minutes before disposal.
l
Do not wet the waste tank cap. If wet, the filter in the cap will cause the tank to malfunction. To keep the
cap dry, place it on the bench label-side up when it is not on the tank.
l
If the waste tank overflows always install a replacement waste tank cap and baffle.