Replacing the waste container cap
Introduction
This topic describes how to replace the waste container cap. Replace the cap once a month.
Contact with biological specimens and materials can transmit potentially fatal disease.
To prevent exposure to biohazardous agents:
l
Put the cytometer in standby mode before disconnecting the waste tank to avoid leakage of biohazardous
waste.
l
The waste container can become pressurized when the cytometer is running. Always disconnect the waste
container from the cytometer before you empty it. Wait at least 30seconds for pressure to dissipate before
you remove the waste cap or sensor.
l
Expose waste container contents to bleach (10% of total volume) for 30minutes before disposal.
Procedure
To replace the cap:
1. Put the cytometer in standby mode.
2. Disconnect the waste container’s sensor line and orange waste line from the waste container tank, and wait
at least 30seconds for pressure to dissipate.
3. Remove the waste cap and attached baffle from the container and place on the bench label-side up.
4. Detach the cap from the baffle. If there is fluid in the baffle, replace it with a new one.
5. Place a new cap on the baffle.
6. Write the date on the cap label.
338677 Rev A
Waste (A)
Space for date
7. Screw the cap assembly onto the waste container and hand-tighten until it is fully closed.
To prevent waste container overpressurization, do not overtighten the trap or attached filter cap. Tighten
each component only until it is hand-tight. Do not use sealants such as Teflon® tape or other adhesives.
Re-attach the alarm sensor line and waste line to the waste container tank.
40 BD FACSymphony™ A5 SE User's Guide