5. Replace the outer sleeve in the retainer.
6. Re-install the retainer and the outer sleeve.
7. Install a sample tube on the SIP to ensure that the outer sleeve has been properly installed.
If the sleeve hits the bottom of the tube, loosen the retainer slightly and push the sleeve up as far as it will
go. Tighten the retainer.
Cleaning or replacing the sheath gasket
Introduction
This topic describes how to clean or replace the gasket of the sheath tank lid.
When to change the sheath filter
We recommend cleaning the sheath gasket when needed.
Procedure
To clean or replace the gasket:
1. Put the cytometer in standby mode.
2. Depressurize the sheath container by pulling up on the vent valve.
3. Remove the lid from the sheath tank.
4. Remove the black gasket from the lid.
5. Rinse it with water to clean the gasket.
6. Place the clean gasket or the new gasket on the lid and make sure the gasket is seated properly on the lid.
Chapter 4 Maintenance 45