104 BD LSR II User’s Guide
5 Open the roller clamp by the fluidics interconnect, and drain the fluid into
a beaker for 5 seconds.
6 Remove the DI water tube from the SIP.
7 Prime twice (perform the following twice):
Press the PRIME button on the fluidics control panel.
When the STNDBY button lights (amber), press the PRIME button
again.
8 Install a tube with 3 mL of a 1:10 dilution of bleach or full-strength
BD FACSClean solution on the SIP.
9 Press the RUN fluid control button; run the cytometer on HI for 30
minutes.
10 Press the STNDBY fluid control button and depressurize the sheath
container by lifting the vent valve.
11 Repeat steps 2 through 10 with BD FACSRinse solution in place of the
bleach solution.
12 Repeat steps 2 through 10 with DI water in place of the BD FACSRinse
solution.
13 Replace the sheath filter and refill the sheath container with sheath fluid.