9-17
TotalAlert Innity™ HTM Medical Gas Notication System
4107 9016 60.01
return key while typing if you wish to move
to the second line of text early.
24. To save, select “Save Low Alarm Instruc-
tions” button.
Figure 299
25. To setup the instructions for a high
pressure fault condition, select the lower
right corner of the badge (location of the
red blinking light when alarm goes in to
fault). Enter non-latin language characters
for the high pressure alarm instructions.
(See Figure 300)
Figure 300
26. To save, select “Save High Alarm Instruc-
tions” button. Touch the Back button to
return to the main label screen. (See Figure
301)
Figure 301
The TotalAlert Innity alarm allows the user
to setup non-latin language characters for
the active alarm screens (both Low and
High), as well as the Information Screen
tabs located on the alarm. Touch the “Other
Labels” tab for setup.
27. To setup the active alarm screen Low,
from the pull down menu, choose “Active
Alarm Screen Message Low”. Then input the
non-latin language character for “LOW” in
the writable tab.
28. To save, select “Save Label” button. (See
Figure 302)
Figure 302
29. To setup the active alarm screen High,
from the pull down menu, choose “Active
Alarm Screen Message High”. Then input
the non-latin language character for “HIGH”
in the writable tab.
30. To save, select “Save Label” button. (See