Transducer and System Maintenance
T3300 Diagnostic Ultrasound System | 96 | B00601-010 4/10/17
WARNING
Do not use a surgeon’s brush when cleaning transducers. Even the use of soft brushes can
damage transducers.
WARNING
Do not use paper products or products that are abrasive when cleaning the transducer. They
damage the soft lens of the acoustic window of the transducer.
WARNING
During cleaning, disinfection, and sterilization, orient the connector and cable that must
remain dry higher than the wet parts, until all parts are dry. This helps keep liquid from
entering unsealed areas of the transducer.
WARNING
When cleaning and disinfecting transducers, do not allow any fluid to enter electrical
connections or metal portions of the connector. Damage due to fluids in these areas is not
covered by the warranty or your service contract.
WARNING
When using an enzymatic cleaner, be sure to use the proper concentration and rinse
thoroughly.
WARNING
Before storing transducers, ensure that they are thoroughly dry. If it is necessary to dry the
transducer lens or acoustic window after cleaning, use a soft cloth and a blotting motion,
instead of a wiping motion.
WARNING
Use only liquid solutions to sterilize transducers. Using autoclave, gas (EtO), or other
methods not approved by BenQ Medical Technology will damage your transducer and void
your warranty.
WARNING
Do not soak the transducer for extended periods of time. Limit the time and depth that
transducers are soaked in disinfectant solution to the minimum time recommended by the
disinfectant manufacturer.
9.4 Inspecting the Transducer
Inspect the transducer’s acoustic lens, the cable, and the transducer connector before each use.
Check the transducer carefully and see if there are cracks, cuts, or any other damages which may
admit fluids. Do not use the transducer if it is damaged.
If the transducer is dropped, examine it immediately for signs of damage. Perform a sample scan to
make sure it operates correctly. Contact BenQ Medical Technology for service or replacement if any
abnormalities are found.