S10/S12/S10A/S12A Patient Monitor User’s Manual
8-28
as the baseline. The baseline value will be used to calculate the QTc value.
After the new QT baseline is set, the original baseline will be discarded. The
baseline will be cleared when the patient is released.
Select Show Baseline or Hide Baseline to show or hide QT baseline
waveform.
CAUTION
Changing QT baseline will affect ΔQTc value and ΔQT alarm.
8.8.6. QT Setting
8.8.6.1. Setting the QT Alarm
QT alarm is set as follows:
1. Select QT parameter area to enter QT menu.
2. Select Alarm submenu.
3. Set properties of QTc and QTc alarm.
8.8.6.2. Selecting QTc Formula
The monitor uses Hodges formula by default to correct QT interval according to heart
rate. If you need to select other QTc formulas, the steps are as follows:
1. Select Main Menu quick keyfrom System column to select
Maintenanceinput maintenance passwordEnter.
2. SelectModulesubmenuECGsubmenu.
3. SelectQTc Formula.
Hodges:
 QTc 1.75 󰇛HeartRate 60󰇜
Bazett: QTc QT 󰇡


󰇢
Fridericia: QTc QT 󰇡


󰇢