S10/S12/S10A/S12A Patient Monitor User’s Manual
17-8
at the top right corner of the event list. For example, 3/10 indicates that there are a total of
10 selected events, and currently there are 3 events.
17.2.7.2. Configuring the Filter
You can filter events by time, alarm priority, parameter category or event type. To
configure the filter, follow this procedure:
1. Enter events review page to switch on 【Filter】.
2. Select 【Filter Setup】 and set the desired filter criterion. Events after
filtering will be displayed in the event list.
CAUTION:
If 【Filter】 is not switch on, the relevant setting in 【Filter Setup】 will not take
effect.
17.2.7.3. Viewing Event Details
To view waveforms and parameter values at the selected event time, follow this
procedure:
1. Enter the event review page.
2. Select 【Detail】.
CAUTION:
Please ensure that the best ECG lead with largest waveform amplitude and the
highest signal-to-noise ration is selected. Choosing the best ECG lead is very
important to recognize cardiac beat, classify cardiac beat and recognize
ventricular fibrillation.
17.2.8. Full Disclosure Review Page
On the Full Disclosure review page, you can review waveform data up to 72 hours.
You can view compressed waveforms, full waveforms and numeric values.