S10/S12/S10A/S12A Patient Monitor User’s Manual
18-2
Early warning scores are not available for pregnant women, those with COPD
(chronic obstructive pulmonary disease), and those under 16 years of age.
18.1.1. Parameters participating in the scoring
The following table lists parameters used for evaluation by each scoring system:
National early warning score (NEWS) Modified early warning score (MEWS)
RR, SpO2, Oxygen Supply, TEMP,
NIBP-Sys, HR/PR, Consciousness
(AVPU)
RR, TEMP, NIBP-Sys, HR/PR,
Consciousness (AVPU)
18.1.2. Display EWS Parameter Area
The steps to display the EWS parameter area are as follows:
1. Enter the Screen layoutpage in one of the following ways:
Select Screen Setupquick key select Screen layout
submenu.
Select Main Menuquick key select Screen layout from
Displaycolumn.
2. Select the parameter area where you want to display EWS, and select
EWS from the pop-up parameter list. The following figure shows an
example of EWS parameter area. The display of EWS parameter area will
vary according to the settings:
(1) Name of scoring system
(2) Total score, the color of the circle indicates the current score level.
(3) This scoring time
2
6
4
5
1
3