S10/S12/S10A/S12A Patient Monitor User’s Manual
18-5
NIBP: The monitor automatically calculates the score after each
NIBP measurement is completed.
Alarm: Scores are automatically calculated after physiological
alarm occurs to scoring parameters.
18.1.6. Setting EWS
18.1.6.1. Select a scoring system
The monitor is equipped with a default scoring system. You can select other
scoring systems as required, as follows:
1. Select the Setupbutton in the scoring interface of EWS.
2. Set Score.
18.1.6.2. Set scoring interval time
When the Interval Modein the Auto Scoringarea is selected, the
measurement interval for automatic scoring can be set as follows:
1. Select theSetupbutton from the scoring interface of EWS.
2. SetInterval.
18.1.6.3. Set the invalid time of the parameter
For manually entered parameter values, you can set the invalid time of the
parameter values as follows:
1. Select the Setup button from the scoring interface of EWS.
2. Set Manual Data Timeout.
18.1.7. EWS score review
In EWS screen, select Tabular Trendssubmenu or Graphicsubmenu to
view the parameter values and scores of all measurement parameters and input
parameters.