S10/S12/S10A/S12A Patient Monitor User’s Manual
18-4
from manual input.
(3) Total score. The color of the circle indicates the current score level.
(4) Indicator of scoring grade: increasing from top to bottom according to the
degree of early warning danger. The current level is shown in the box.
(5) This scoring time
(6) Countdown to next scoring
(7) Scoring interval
(8) Recommended action measures
18.1.4. Calculation of Scores
Calculate the score as follows:
1. Select Resetto clear the last scoring result, and refresh the parameter
values automatically obtained from the monitor and the scoring of
corresponding parameters.
2. Measure or manually enter parameter values of other parameters.
3. Select Calculate to obtain the scoring results.
CAUTION:
Before each scoring, please press the Reset key to clear the last scoring result.
The keyboard symbol to the right of the parameter value indicates that the
parameter value is manually entered.
You can calculate the score only when the parameter values of all the parameters
involved in the calculation are valid.
18.1.5. Automatic Scoring
Set up the automatic scoring method as follows:
1. Select the Setup button from the scoring interface of EWS.
2. In the Auto Scoringarea, select as required:
Interval Mode: The monitor automatically calculates the score
according to the set time interval.