S10/S12/S10A/S12A Patient Monitor User’s Manual
18-9
Severe 3-4 Red Can be brain death or remain
vegetative.
18.2.4. Set GCS score invalid time
Select the 【Invalid Time】 button in the GCS menu to set the invalid time of the
GCS score. If the set score interval is reached without re-scoring, the original score is
invalidated and displayed in a hollow word.
18.2.5. Set the GCS score threshold
The coma score threshold was set as follows:
1. Select 【 Main Menu】quick key → select 【Maintenance】 from
【System】column→input maintenance password→enter.
2. Select 【CAA】→【GCS】submenu.
3. Set the high and low limits of 【Mild】, 【Moderate】, and 【Severe】
respectively.
18.2.6. GCS score review
In the GCS menu, select the 【Review】button to enter the 【Review】 menu, and
view the GCS score trend from the 【Trend Table】page.