S10/S12/S10A/S12A Patient Monitor User’s Manual
19-5
19.4.2. Input Parameters
Abbreviation Unit Full Name
C.O. L/min cardiac output
HR bpm heart rate
PAWP mmHg pulmonary artery wedge pressure
MAP mmHg artery mean pressure
MPAP mmHg mean pulmonary artery pressure
CVP mmHg central venous pressure
EDV mL end-diastolic volume
Height cm height
Weight kg weight
19.4.3. Output Parameters and calculation formula
Output
Parameter
Unit Full Name Formula
C.I. mL/min/m
2
cardiac index
C.I. C.O. / BSA
BSA m
2
body surface
area
BSA
HT
.
WT
.
0.007184
SV mL stroke volume SV = 1000× C.O. /HR
SVI mL/m
2
stroke index SVI= SV/BSA
SVR dyn*s/cm
5
systemic
vascular
resistance
SV
R
79.96
MAP CVP
C.O.
SVRI
dyn*s*m
2
/c
m
5
systemic
vascular
resistance index
SVRI = SVI /BSA

Table of Contents