EasyManua.ls Logo

Biolight Meditech S10 - Page 201

Default Icon
307 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
S10/S12/S10A/S12A Patient Monitor User’s Manual
19-6
Output
Parameter
Unit Full Name Formula
PVR dyn*s /cm
5
pulmonary
vascular
resistance
PV
R
79.96
MPAP PAWP
C.O.
PVRI
dyn*s*m
2
/c
m
5
pulmonary
vascular
resistance index
PVRI= PVR×BSA
LCW kg*m left cardiac work LCW= 0.0136 × APMAP × C.O.
LCWI kg*m/m
2
left cardiac work
index
LCWI = RCW×BSA
LVSW g*m
left
ventricularstrok
e work
LVSW = 0.0136 ×MAP×SV
LVSWI g*m/m
2
left ventricular
stroke work
index
LVSWI = LVSW/BSA
RCW kg*m
right cardiac
work
RC
W
0.0136 PAMAP
C.O.
RCWI kg*m/m
2
right cardiac
work index
RCWI= RCW/BSA
RVSW g*m
right ventricular
stroke work
RVSW = 0.0136 × MPAP× SV
RVSWI g*m/m
2
right ventricular
stroke work
index
R VSWI= RVSW /BSA
EF % ejection fraction EF= 100 × SV /EDV
19.5. Oxygenation Calculation
The monitor provides the oxygenation calculation function. The monitor can save
the results of up to 20 calculations, which are displayed in groups:

Table of Contents