S10/S12/S10A/S12A Patient Monitor User’s Manual
15
A.8.3.2 Measurement Specification
Measurement
parameters
SYS, DIA, MAP,PR
Mode of operation
Manual, Auto, STAT, Sequence
Intervals for periodic
measurement time
1min, 2 min, 2.5 min, 3 min, 5 min, 10 min, 15 min, 20 min,
30 min, 1h, 1.5h, 2h, 3h, 4h, 8h
Sequence mode
Up to 5 groups are supported, and each group individually sets
the interval and number of periodic measurement.
STAT mode cycle
time
5 min
Measurement range
of cuff pressure
0~300 mmHg
Initial inflation
pressure
Adult: 100~280mmHg, default 160mmHg
Pediatric: 100~240 mmHg, default 130mmHg
Neonatal: 60~140mmHg, default 100mmHg
Sensor calibration
time
One year (recommend)
Unit
mmHg, kPa
Dynamic pressure
measurement range
Systolic
Adult: 30~270 mmHg
Pediatric: 30~235 mmHg
Neonatal: 30~135 mmHg
Diastolic
Adult: 10~220 mmHg
Pediatric: 10~220 mmHg
Neonatal: 10~110 mmHg
Mean
Adult: 20~235 mmHg
Pediatric: 20~225 mmHg
Neonatal: 20~125 mmHg
Dynamic Pressure
Measurement Error
of Simulator
Systolic 35~255mmHg: ±8 mmHg
Diastolic 15~195mmHg: ±8 mmHg
Mean 22~215mmHg: ±8 mmHg
Static pressure
±3 mmHg(±0.4kPa)