S10/S12/S10A/S12A Patient Monitor User’s Manual
19
Accuracy of
circuit
±0.1
Power supplied to probe
<20W
Updated time
Every about 1~2s
Minimum measurement
time
Surface: 100s
Coelom: 80s
A.8.6 IBP
A.8.6.1 Standard
Meet the standard of IEC 60601-2-34.
A.8.6.2 Functional specification
Measurement
parameters
Dual channels IBP parameters (including systolic blood
pressure, diastolic blood pressure, average pressure, PR) and
waveforms
Measurement site
ART/CVP/ICP/PA/Ao/UAP/BAP/FAP//LAP/RAP/UVP/LV/P
AWP, additionally, P1 and P2 are arbitrary sites
Scale
Manual, interval and automatic scale settings
Unit
mmHg, kPa, cmH2O
PPV
Measurement
range
0%~50%
Resolution 1%
Static pressure
measurement range
and accuracy
Measurement
range
-50 mmHg ~+360mmHg
Resolution 1 mmHg
Accuracy
±2mmHg or ±2% (whichever is greater,
without sensor)
Dynamic pressure
measurement range
Measurement
range
-50 mmHg ~+360mmHg