S10/S12/S10A/S12A Patient Monitor User’s Manual
20
and accuracy
Accuracy
±2mmHg or ±2% (whichever is greater,
without sensor)
Frequency response
Including sensor At least d.c.~10Hz
Only host At least d.c.~12Hz
IBP zero range
-200mmHg+200mmHg
PR
Measurement
range
30bpm ~300bpm
Resolution 1bpm
Accuracy ±1% or ±1bpm whichever is greater
Pressure sensor
Nominal
sensitivity
5uV/V/ mmHg
Output
impedance
300~3000
Volumetric
displacement
<0.04 mm
3
/100 mmHg
Error ±2%
IBP analog output
delay
35ms
A.8.7 CO2
A.8.7.1 Standard
Meet the standard of ISO 80601-2-55.
A.8.7.2 Functional Specification
Measurement
parameter
EtCO2, FiCO2, a CO2 waveform and awRR
Measurement
method
Mainstream, Sidestream