S10/S12/S10A/S12A Patient Monitor User’s Manual
21
Unit
mmHg, kPa and %
A.8.7.3 Performance Specification
EtCO2/FiCO2
measurement range
BLT Capno_S
0%19.7% (0mmHg150mmHg)
Respironics
LOFLO C5
0%19.7% (0mmHg150mmHg)
BLT Capno_M
0%19.7% (0mmHg150mmHg)
Respironics
CAPNOSTA 5
0%19.7% (0mmHg150mmHg)
Masimo ISA
Capno
0%25% (0mmHg190mmHg)
Masimo IRMA
0%25% (0mmHg190mmHg)
EtCO2/FiCO2
measurement
accuracy
± (0.43% + 8% of reading)
Preheating time
BLT Capno_S
97% of the design error can be reached
within 45s, and the design error can be
reached within 2 min.
Respironics
LOFLO C5
<20s
Masimo ISA
Capno
<10s (Report concentration and achieve
highest accuracy)
BLT Capno_M
97% of the design error can be reached
within 8s, and the design error can be
reached within 20s.
Respironics
CAPNOSTA 5
<15s
Masimo IRMA <10s
EtCO2/FiCO2 display
resolution
0.1% or 1mmHg