S10/S12/S10A/S12A Patient Monitor User’s Manual
22
awRR measurement
range
BLT Capno_S
3150 bpm
Respironics
LOFLO C5
2150 bpm
Masimo ISA
Capno
0150 bpm
BLT Capno_M
3150 bpm
Respironics
CAPNOSTA 5
0150 bpm
Masimo IRMA
0150 bpm
awRR measurement
accuracy
±1 bpm
Sampling frequency
and accuracy of gas
(only sidestream)
BLT Capno_S 50±10mL/min
Respironics
LOFLO C5
50±10mL/min
Masimo ISA
Capno
50±10mL/min
Response time
BLT Capno_S <3s (including delay time and rise time)
Respironics
LOFLO C5
<3s (including delay time and rise time)
Masimo ISA
Capno
<3s (including delay time and rise time)
BLT Capno_M About 70ms (rising time)
Respironics
CAPNOSTA 5
<60ms (rising time)
Masimo IRMA <90ms (rising time)
A.8.7.4 The effects on CO2 measuring values caused by the interfering gases
BLT Capno_S & BLT Capno_M
The accuracy of CO2 is affected by interfering gases and water vapor. For example,
N2O, a halide-containing anesthetic gas can raise the CO2 reading (about 2%-10%), and