S10/S12/S10A/S12A Patient Monitor User’s Manual
23
helium and oxygen can reduce the CO2 reading (1%-10%), so in the presence of
interfering gas, the user should send relevant command to the module (the instrument’s
compensation menu to adjust the interference gas data), so that the module (instrument)
can meet the nominal accuracy requirements.
Respironics LOFLO C5
Gas or Vapor Gas Level Quantitative Effects
Nitrous oxide 60% No Additional Effect
Halothane 4% No Additional Effect
Enflurane 5% No Additional Effect
Isoflurane 5% No Additional Effect
Sevoflurane 5% No Additional Effect
Xenon 80% Negatively bias Carbon Dioxide
values by up to an additional 5
mmHg at 38 mmHg
Helium 50% No Additional Effect
Metered dose inhaler
propellants
Unspecified Unspecified
Desflurane 15% Concentrations greater than 5% will
positively bias Carbon Dioxide
values by up to an additional 3
mmHg at 38 mmHg.
Ethanol 0.1% No Additional Effect
Isopropanol 0.1% No Additional Effect
Acetone 0.1% No Additional Effect
Methane 1% No Additional Effect
Masimo ISA Capno & Masimo IRMA
Gas Gas level CO2 Agents N2O
ISA CO2 ISA AX+
ISA OR+
N2O
4
)
60 vol%
2
)
1
)
1
)
1
)