S10/S12/S10A/S12A Patient Monitor User’s Manual
24
HAL
4
4 vol% –
1
–
1
–
1
–
1
ENF, ISO, SEV
4
5 vol% +8% of
reading
3)
–
1
–
1
–
1
DES
4
15 vol% +12% of
reading
3)
–
1
–
1
–
1
Xe (Xenon)
4
80 vol% 10% of reading
3
–
1
–
1
He (Helium)
4
50 vol% 6% of reading
3
–
1
–
1
Metered dose
inhaler
propellants
4)
Not for use with metered dose inhaler propellants
C2H5OH
(Ethanol)
4)
0.3 vol% –
1
–
1
–
1
–
1
C3H7OH
(Isopropanol)
4)
0.5 vol% –
1
–
1
–
1
–
1
CH3COCH3
(Acetone)
4)
1 vol% –
1
–
1
–
1
–
1
CH4 (Methane)
4)
3 vol% –
1
–
1
–
1
–
1
CO (Carbon
monoxide)
5)
1 vol% –
1
–
1
–
1
–
1
NO (Nitrogen
monoxide)
5)
0.02 vol% –
1
–
1
–
1
–
1
O2
5
100 vol% –
2
–
2
–
1
Note 1: Negligible interference, effect included in the specification “Accuracy, all
conditions” above.
Note 2: Negligible interference with N2O / O2 concentrations correctly set, effect
included in the specification “Accuracy, all conditions” above.
Note 3: Interference at indicated gas level. For example, 50 vol% Helium typically
decreases the CO2 readings by 6%. This means that if measuring on a mixture containing
5.0 vol% CO2 and 50 vol% Helium, the actual measured CO2 concentration will typically
be (1-0.06) * 5.0 vol% = 4.7 vol% CO2.
Note 4: According to the EN ISO 80601-2-55:2011 standard.