S10/S12/S10A/S12A Patient Monitor User’s Manual
3-13
3.7.5. Privacy Mode
The privacy mode is a special monitoring mode. In the privacy mode, the monitor
does not display patient information and monitoring data to protect patient information
from non-clinicians such as visitors.
The privacy mode is only available when the patient admitted by the monitor is also
monitored by the Central monitoring system. The monitor continues monitoring the
patient, but patient data is only visible at the Central monitoring system.
3.7.5.1. Entering the Privacy Mode
To enter the privacy mode, choose either of the following ways:
Select Main Menu quick keyfrom theDisplaycolumn
to select
Privacy ModebuttonSelectOK.
The monitor has the following features after entering the privacy mode:
The screen turns blank, and prompt Being monitoringat same time.
All parameters and waveforms display are shield.
Except for the low battery alarm, the monitor inactivates alarm tone and alarm
light of all other alarms.
The monitor suppresses all system prompt tone, including heart beat tone,
pulse tone, etc.
WARNING
In Privacy mode, all audible alarms are suppressed and the alarm light is
deactivated at the monitor. Alarms are presented only at the
Central
monitoring system. Pay attention to potential risk.
CAUTION
Cannot enter the privacy mode if a low battery alarm occurs.