5.1.3. Dispensing system
5.1.3.1. Changing the thermostated pipe.
a) Remove the arm casing.
b) It is recommendable to remove the needle before handling the unit to prevent it from being damaged.
c) Disconnect the electrical connector (1) and the Teon pipe (2) and remove the anges (3).
d) Remove the protective cover (4).
e) Remove the entire thermostated pipe by removing the 3 screws (5) that hold it in place.
f) Ret the new unit, replacing the anges.