EasyManuals Logo

biosystems A15 User Manual

biosystems A15
131 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #67 background imageLoading...
Page #67 background image
67
4.2.6. Adjustment of the level control scales
This screen makes it possible to set the level control scales with the empty waste and distilled water containers
(0% capacity) and when they are full (100% capacity). The maximum capacity of the containers is approximately
3L. The technician must choose whether he wishes to set the distilled water or waste container scales, with the
corresponding container full or empty. According to the requested adjustment, the corresponding container, full
or empty, must be placed in position and the Adjust button pressed. Based on the settings made, the analyzer
automatically adjusts the scales. On pressing the Store button, the analyzer saves the new values of the adjus-
ted parameters. Pressing the Cancel button keeps the last stored values and the current values are not stored.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the biosystems A15 and is the answer not in the manual?

biosystems A15 Specifications

General IconGeneral
Brandbiosystems
ModelA15
CategoryLaboratory Equipment
LanguageEnglish