4.3.5. Needle thermostatation system test
This screen makes it possible to check that the dispensing temperature of the reactions is around 37ºC. To make
this adjustment, the analyzer must be initialised. The technician must measure the temperature of the dispen-
sed liquid with a thermometer calibrated at 37ºC. The program shows the set point temperature of the current
control. This parameter must be different from 37ºC. When the technician so indicates, the analyzer dispenses
thermostated distilled water on a certain position in the racks tray shown on the screen. The technician must
measure the temperature of the water with the calibrated thermometer and introduce the temperature on the
screen. The program indicates if the temperature measured is within the tolerated error margins and stores this
value for the test result reports. The liquid to be dispensed is taken from the system liquid container or from the
bottle of reagent selected by the technician.